Patient results and laboratory test accuracy

Alexander Katayev (Department of Science and Technology, Laboratory Corporation of America Holdings, Elon, North Carolina, USA)
James K. Fleming (Department of Science and Technology, Laboratory Corporation of America Holdings, Elon, North Carolina, USA)

International Journal of Health Care Quality Assurance

ISSN: 0952-6862

Article publication date: 4 February 2014

270

Abstract

Purpose

Traditional quality control materials used for monitoring the clinical laboratory test accuracy might be non-commutable with patient samples and may not detect systematic errors. The aim of this paper is to describe a method to monitor inter-instrument bias using result distributions that are independent of the control's commutability.

Design/methodology/approach

Serum calcium data collected within a laboratory network were assessed. A reference interval was calculated using a computerized, indirect Hoffmann's algorithm using all data across a laboratory network without excluding any results. Results outside the reference interval were considered as the zero-bias distribution. Three allowable bias levels were then calculated to determine the corresponding shift in abnormal results for each bias level in both directions from the zero-bias distribution. The observed result distributions in three laboratories within the network were compared for bias performance after one year of the reference interval study.

Findings

Performance levels for bias were: minimum allowable <1.27 percent; desirable <0.85 percent; and optimal <0.42 percent. Zero bias result distribution above and below the reference interval for calcium was 3.92 percent and 2.53 percent respectively. All three laboratories performed within the desirable allowable bias level.

Originality/value

Bias-monitoring process using patient result distributions allows managers to: assess systematic error between laboratory instruments; improve laboratory quality control; and strengthen patient risk management.